Introduction

The fan filter must be replaced every month or as often as required, and the air and HEPA filter may be replaced yearly or as necessary.

Each filter in this guide may be replaced independently from the others, but the instructions have been integrated into one overarching guide as the preventative maintenance for the HAMILTON-C1. Refer to the individual replacement guides of the air and fan filter by clicking the links below:

  • Air FilterFan Filter

Be sure to disconnect the HAMILTON-C1 from any external sources of power before beginning.

What you need

Step 1

              Air Filter               
  • Turn the Hamilton-C1 so that the rear view of the device is facing you.
  • Use your fingers to grasp the current air filter located at number 4 of the graphic and pull the filter out of the slot.

Turn the Hamilton-C1 so that the rear view of the device is facing you.

Use your fingers to grasp the current air filter located at number 4 of the graphic and pull the filter out of the slot.

1024

Step 2

  • Insert a new air filter into the slot located at number 4.

Insert a new air filter into the slot located at number 4.

Step 3

              Fan Filter               
  • Orient the Hamilton-C1 to the right side so that the gas connections are visible.
  • Grab and remove the fan filter located at location number 5 in the graphic.

Orient the Hamilton-C1 to the right side so that the gas connections are visible.

Grab and remove the fan filter located at location number 5 in the graphic.

Step 4

  • Place a new fan filter into the designated slot.

Place a new fan filter into the designated slot.

Step 5

              HEPA Filter               
  • Orient the Hamilton-C1 so the back is facing you.
  • Remove the back panel as labeled by number 6 of the graphic using a Phillips #1 screwdriver.

Orient the Hamilton-C1 so the back is facing you.

Remove the back panel as labeled by number 6 of the graphic using a Phillips #1 screwdriver.

Step 6

  • Remove the current HEPA filter by pulling the filter outward and down.

Remove the current HEPA filter by pulling the filter outward and down.

Step 7

  • Insert the new HEPA filter into the slot.

Insert the new HEPA filter into the slot.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 1 other contributor 

                    Tina Vo                     

Member since: 04/06/2020

538 Reputation

                                      4 Guides authored                  



                       Badges:
                       6







                                                        +3 more badges                           

Team

                       Cal Poly, Team S7-G11, Paton Spring 2020                        

                                                  Member of Cal Poly, Team S7-G11, Paton Spring 2020 



                    CPSU-PATON-S20S7G11                     


                                            2 Members                     


                                            6 Guides authored