Introduction
What you need
Step 1
Opening the Column in the Hamilton Galileo
- Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.
Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.
1024
Step 2
- Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).
Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).
Step 3
- Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.
Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.
Step 4
- Without removing any electrical connections, open the left side of the column as wide as practical.
- This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.
Without removing any electrical connections, open the left side of the column as wide as practical.
This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.
Step 5
- Make sure no electrical connections are touching the column.
- Make sure all green/brown grounding (earthing) cables are securely in place.
Make sure no electrical connections are touching the column.
Make sure all green/brown grounding (earthing) cables are securely in place.
Step 6
Control Board and GMP Assembly
- Locate the control board and GMP assembly.
Locate the control board and GMP assembly.
Step 7
- Remove all cable connectors joining the control board and the GMP board to other parts of the GALILEO.
Remove all cable connectors joining the control board and the GMP board to other parts of the GALILEO.
Step 8
- Remove the single grounding (earthing) screw securing the control board to the aluminium mounting plate.
Remove the single grounding (earthing) screw securing the control board to the aluminium mounting plate.
Step 9
- Remove the control board from the plastic spacers. Do do this, squeeze each spacer in turn, while easing the board off the spacer. Leave the spacer in the center of the board until last.
Remove the control board from the plastic spacers. Do do this, squeeze each spacer in turn, while easing the board off the spacer. Leave the spacer in the center of the board until last.
Step 10
- To re-mount the control board, reverse steps seen here. Only press the control board onto plastic spacers, do not use pliers. See image for correct cable layout.
To re-mount the control board, reverse steps seen here. Only press the control board onto plastic spacers, do not use pliers. See image for correct cable layout.
To reassemble, mount the control board back onto plastic spacers without using pliers and be sure to connect all cables together correctly.
Cancel: I did not complete this guide.
Author
with 1 other contributor
Jacob Dixon
Member since: 04/09/2020
191 Reputation
1 Guide authored
Badges:
4
+1 more badge
Team
Cal Poly, Team S7-G26, Paton Spring 2020
Member of Cal Poly, Team S7-G26, Paton Spring 2020
CPSU-PATON-S20S7G26
3 Members
4 Guides authored