BackHamilton Galileo
Full Screen
Options
History
Save to Favorites
Download PDF
Edit
Translate
Get Shareable Link
Embed This Guide
Notify Me of Changes
Stop Notifications
Introduction
What you need
Step 1
LCD
- First open the main enclosure by removing the four hexagonal-drive (Allen) screws at the rear of the enclosure.
First open the main enclosure by removing the four hexagonal-drive (Allen) screws at the rear of the enclosure.
1024
Step 2
- Unscrew the control board mounting plate.
Unscrew the control board mounting plate.
Step 3
- “Hinge” back the mounting plate, complete with control board. This exposes the LCD.
“Hinge” back the mounting plate, complete with control board. This exposes the LCD.
Step 4
- Unplug the cable from the GMP adaptor board from the LCD.
Unplug the cable from the GMP adaptor board from the LCD.
Step 5
- Unplug from the dc/ac board, the cable to the interface board.
Unplug from the dc/ac board, the cable to the interface board.
Step 6
- Unscrew the four hexagonal-drive (Allen) screws securing the two brackets on which the LCD is mounted.
Unscrew the four hexagonal-drive (Allen) screws securing the two brackets on which the LCD is mounted.
Step 7
- Lift out the complete assembly.
Lift out the complete assembly.
Step 8
- Unplug both of the lights from the dc/ac board.
Unplug both of the lights from the dc/ac board.
Step 9
- Remove the LCD from the two brackets, by removing the black hexagonal-drive (Allen) screws on the LCD. NOTE: Do not remove the silver hexagonal-drive (Allen) screws on the SIDE of the LCD.
Remove the LCD from the two brackets, by removing the black hexagonal-drive (Allen) screws on the LCD. NOTE: Do not remove the silver hexagonal-drive (Allen) screws on the SIDE of the LCD.
Step 10
- When fitting a new display, there must be an offset of approximately 2 mm between the plastic sealing strips on the LCD, and the brackets.
When fitting a new display, there must be an offset of approximately 2 mm between the plastic sealing strips on the LCD, and the brackets.
Step 11
- If necessary, adjust this gap using the silver screws on the side of the LCD.
If necessary, adjust this gap using the silver screws on the side of the LCD.
Step 12
- The surface of the LCD must be clean. Wipe it free of finger marks and dust with an antistatic paper towel.
The surface of the LCD must be clean. Wipe it free of finger marks and dust with an antistatic paper towel.
Step 13
- The cable from the GMP to the LCD must be in the cable grip on the control board mounting plate.
The cable from the GMP to the LCD must be in the cable grip on the control board mounting plate.
Step 14
- When screwing the LCD assembly into GALILEO, you must align it with the glass screen in GALILEO. Check this by looking at the front of GALILEO.
- Turn on GALILEO, and make sure that the LCD lights up, and is properly aligned.
When screwing the LCD assembly into GALILEO, you must align it with the glass screen in GALILEO. Check this by looking at the front of GALILEO.
Turn on GALILEO, and make sure that the LCD lights up, and is properly aligned.
To reassemble your device, follow these instructions in reverse order.