- How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff
- 4 steps
Flags
0
BackFisher Paykel Neopuff/Perivent Infant T-Piece Resuscitator
Full Screen
Options
History
Save to Favorites
Download PDF
Edit
Translate
Get Shareable Link
Embed This Guide
Notify Me of Changes
Stop Notifications
Introduction
Step 1
How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff
- Adjust the gas flow to 10 L/min.
Adjust the gas flow to 10 L/min.
1024
Step 2
- Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.
Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.
Step 3
- Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20
Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20
Step 4
- Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.
Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.
Alternate settings for the Max Pressure Relief should be made in accordance with hospital protocol.