Introduction

For more information on how to perform a circuit check, please refer to page 6-2 in the service manual.

Step 1

              Setup               
  • Connect the breathing circuit to the main flow outlet, or as if it were being used for a patient.
  • Cover up the breathing circuit (do not use test lung) .

Connect the breathing circuit to the main flow outlet, or as if it were being used for a patient.

Cover up the breathing circuit (do not use test lung) .

1024

Step 2

              Begin Test               
  • Touch the Circuit Check button at the top of the screen.
  • Press ‘Accept’.

Touch the Circuit Check button at the top of the screen.

Press ‘Accept’.

Step 3

              Finish               
  • Uncover the end of the patient circuit pictured in Step 1.
  • Press the ‘Accept’ button to continue with the circuit check.
  • If test passes, the message “Circuit Check PASSED press accept to confirm” will appear on the screen.

Uncover the end of the patient circuit pictured in Step 1.

Press the ‘Accept’ button to continue with the circuit check.

If test passes, the message “Circuit Check PASSED press accept to confirm” will appear on the screen.

Step 4

              Warnings/Implications               
  • If the Circuit check does not pass, do not use the ventilator for patient use. Use an alternate Method of ventilation.
  • If the Circuit Check has failed, there may be an issue with the pump assembly, refer to the “Pump Assembly Replacement” guide to proceed with this diagnosis.

If the Circuit check does not pass, do not use the ventilator for patient use. Use an alternate Method of ventilation.

If the Circuit Check has failed, there may be an issue with the pump assembly, refer to the “Pump Assembly Replacement” guide to proceed with this diagnosis.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 2 other contributors 

                    Merek McGill                     

Member since: 04/07/2020

520 Reputation

                                      5 Guides authored                  



                       Badges:
                       7







                                                        +4 more badges                           

Team

                       Cal Poly, Team S7-G5, Paton Spring 2020                        

                                                  Member of Cal Poly, Team S7-G5, Paton Spring 2020 



                    CPSU-PATON-S20S7G5                     


                                            3 Members                     


                                            6 Guides authored