Introduction

Step 1

              How to Test the Valve System on a Fisher Paykel Neopuff               
  • Set the gas supply to 5 L/min.

Set the gas supply to 5 L/min.

1024

Step 2

  • Completely close the Peak Inspiratory Pressure control and the Max Pressure control by turning both knobs completely clockwise.

Completely close the Peak Inspiratory Pressure control and the Max Pressure control by turning both knobs completely clockwise.

Step 3

  • Close the bleed valve and check that the gauge reads at least 60 cmh2O.

Close the bleed valve and check that the gauge reads at least 60 cmh2O.

Step 4

  • Set the gas supply to 15 L/min.

Set the gas supply to 15 L/min.

Step 5

  • Close the bleed valve and check that the gauge reads no higher than 80 cmH2O.

Close the bleed valve and check that the gauge reads no higher than 80 cmH2O.

Step 6

  • Set the gas supply to 10 L/min.

Set the gas supply to 10 L/min.

Step 7

  • Close the bleed valve and turn the Max Pressure Relief valve until the manometer reads 40 cmh2O.

Close the bleed valve and turn the Max Pressure Relief valve until the manometer reads 40 cmh2O.

Step 8

  • Check that the manometer needle rises and falls smoothly.

Check that the manometer needle rises and falls smoothly.

Step 9

  • Gently rotate the Max Pressure Relief cover over the Max Pressure Relief knob.

Gently rotate the Max Pressure Relief cover over the Max Pressure Relief knob.

Step 10

  • Reset the peak inspiratory pressure to 20 cmh2O and turn off the gas flow. Testing is now complete. If the F&P Neopuff/Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers).

Reset the peak inspiratory pressure to 20 cmh2O and turn off the gas flow. Testing is now complete. If the F&P Neopuff/Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers).

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 2 other contributors 

                    Constantine Yap                     

Member since: 05/04/2020

337 Reputation

                                      2 Guides authored                  



                       Badges:
                       3

Team

                       Cal Poly, Team S1-G6, Zimmerman Spring 2020                        

                                                  Member of Cal Poly, Team S1-G6, Zimmerman Spring 2020 



                    CPSU-ZIMMERMAN-S20S1G6                     


                                            4 Members                     


                                            26 Guides authored